System is communicating System is communicating #1: My chromato- gram is empty MASS SPECTROMETER (MS) TROUBLESHOOTING B – Verify accuracy of MS method: Ensure the MS method is properly encoded to acquire the desired data and required reagents are prepared. Solutions Missing a solve for troubleshooting your instrument? Contact ptfionbaording@gmail.com A – Inspect ESI capillary for clog and review source settings: Irregular and/or absent spray can be caused by a clog. Follow instrument manufacturer instructions to check for clogs and adjust source settings to ensure consistent spray. C – Examine LC-MS for leaks and/or clogs: Investigate irregular pressure using an LC troubleshooting guide. D – Check system for contamination: Move to issue #3: my blank files have a lot of signal. *Some instrument allow you to remove the nebulizer to check spray performance. Others have windows into the source to view spray performance. Are any mass signals present? Yes Check the system pressure at starting conditions Pressure is Low or High Pressure is Normal See Solution ANo/Unsure Spray is absent or sputtering Spray is consistent See Solution B *Check the stability of the ionization spray See Solution D See Solution C #2: My observed mass values are shifted or inaccurate #3: My blank files have a lot of signal F – Re-calibrate mass spectrometer: Follow instrument manufacturer instructions to calibrate instrument (see Note 1). Did you verify MS calibration before acquisition? No/Unsure No See Solution F Yes Are you acquiring any reference mass data? Is this vial full and correctly applied in the MS method? No/Unsure Fill vial and see Solution BYes No/Unsure Are low mass ions most affected? Yes/Unsure See Solution E E – Perform check on detector voltage: Follow instrument manufacturer instructions to perform detector voltage check. Yes #4: ***The instrument has stopped communicating! Do MS peaks have normal distributions? Notes Note 1: Rebooting computer and/or mass spectrometer can help improve calibration application. Yes/Unsure No See Solution G G – Re-tune mass spectrometer: A tune of mass spectrometer voltages can help ensure correct mass value assignments. Does the signal in the blanks appear to be related to any injections preceding the blank? No/Unsure Yes Reboot computer connected to the MS and acquisition software ***Instrument manufacturers have specific processes for shutting down mass spectrometers. See Solution H H – Use stronger needle wash: Needle washes that clean the autosampler between injections can help reduce carryover. Alternatively, consider running additional blanks between samples. Has column been cleaned and was it stored properly? No/Unsure Yes **Run system with 100%B then 100%A through column **To avoid dirtying the mass spectrometer, send LC to waste during wash. Signal persists I – Re-make mobile phases: Follow the mobile phase making procedure outlined in Reverse Phase Metabolomics SOPs. Ensure all reagents are LC-MS grade, or equivalent. See Solution I System is NOT communicating Follow instrument vendor guidelines for soft system reboot System is NOT communicating Follow instrument vendor guidelines for hard system reboot J – Reduce extra programs on MS computer: Try to reduce activities not related to acquiring MS data to reduce reboot requirements (see Note 2). Note 2: It is recommended to recalibrate MS after every reboot and re-start data acquisitions. See Solution J See Solution J Slide 2